CVE-2014-5434

CRITICAL

Baxter SIGMA Spectrum Infusion System <6.05 - Info Disclosure

Title source: llm
STIX 2.1

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Scores

CVSS v3 9.8
EPSS 0.0025
EPSS Percentile 48.0%
Attack Vector NETWORK
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Details

CWE
CWE-259 CWE-798
Status published
Products (1)
baxter/sigma_spectrum_infusion_system_firmware 6.05
Published Mar 26, 2019
Tracked Since Feb 18, 2026