CVE-2015-1012

HIGH

Hospira LifeCare PCA Infusion System <5 - Info Disclosure

Title source: llm
STIX 2.1

Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

References (1)

Core 1
Core References
Mitigation, Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01

Scores

CVSS v3 7.5
EPSS 0.0077
EPSS Percentile 50.8%
Attack Vector NETWORK
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Details

CWE
CWE-200 CWE-312
Status published
Products (1)
pfizer/lifecare_pca_infusion_system_firmware < 5.0
Published Mar 25, 2019
Tracked Since Feb 18, 2026