CVE-2015-3954

CRITICAL

Hospira <13.4 - Privilege Escalation

Title source: llm
STIX 2.1

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

References (1)

Core 1
Core References
Mitigation, Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01

Scores

CVSS v3 9.8
EPSS 0.0204
EPSS Percentile 78.6%
Attack Vector NETWORK
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Details

CWE
CWE-285
Status published
Products (3)
pifzer/plum_a\+3_infusion_system_firmware < 13.6
pifzer/plum_a\+_infusion_system_firmware < 13.4
pifzer/symbiq_infusion_system_firmware < 3.13
Published Mar 25, 2019
Tracked Since Feb 18, 2026