CVE-2015-3956

CRITICAL

Hospira <13.4 - Unauthenticated RCE

Title source: llm
STIX 2.1

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

References (1)

Core 1
Core References
Mitigation, Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01

Scores

CVSS v3 9.8
EPSS 0.0095
EPSS Percentile 56.6%
Attack Vector NETWORK
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Details

CWE
CWE-345
Status published
Products (3)
pifzer/plum_a\+3_infusion_system_firmware < 13.6
pifzer/plum_a\+_infusion_system_firmware < 13.4
pifzer/symbiq_infusion_system_firmware < 3.13
Published Mar 25, 2019
Tracked Since Feb 18, 2026