CVE-2017-11578
MEDIUMBlipcare Wi-Fi Blood Pressure Monitor < bp700_10.1 - Sensitive Information Exposure via Plaintext HTTP
Title source: llmDescription
It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.
References (3)
Core 3
Core References
Mailing List, Third Party Advisory mailing-list
x_refsource_bugtraq
https://seclists.org/bugtraq/2019/Jun/8
Third Party Advisory, VDB Entry x_refsource_misc
http://packetstormsecurity.com/files/153225/Blipcare-Clear-Text-Communication-Memory-Corruption.html
Exploit, Third Party Advisory x_refsource_misc
https://github.com/ethanhunnt/IoT_vulnerabilities/blob/master/Blipcare_sec_issues.pdf
Scores
CVSS v3
5.9
EPSS
0.0205
EPSS Percentile
78.9%
Attack Vector
NETWORK
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
Details
CWE
CWE-200
Status
published
Products (1)
blipcare/wi-fi_blood_pressure_monitor_firmware
< bp700_10.1
Published
Jul 02, 2019
Tracked Since
Feb 18, 2026