CVE-2017-12725

MEDIUM

Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1, 1.5, 1.6 - Use of Hard-coded Credentials

Title source: llm
STIX 2.1

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

References (2)

Core 2
Core References
Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A
Third Party Advisory, VDB Entry vdb-entry x_refsource_bid
http://www.securityfocus.com/bid/100665

Scores

CVSS v3 5.6
EPSS 0.0086
EPSS Percentile 53.8%
Attack Vector NETWORK
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L

Details

CWE
CWE-798
Status published
Products (3)
smiths-medical/medfusion_4000_wireless_syringe_infusion_pump 1.1
smiths-medical/medfusion_4000_wireless_syringe_infusion_pump 1.5
smiths-medical/medfusion_4000_wireless_syringe_infusion_pump 1.6
Published Feb 15, 2018
Tracked Since Feb 18, 2026