CVE-2017-12725
MEDIUMSmiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1, 1.5, 1.6 - Use of Hard-coded Credentials
Title source: llmDescription
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
References (2)
Core 2
Core References
Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A
Third Party Advisory, VDB Entry vdb-entry
x_refsource_bid
http://www.securityfocus.com/bid/100665
Scores
CVSS v3
5.6
EPSS
0.0086
EPSS Percentile
53.8%
Attack Vector
NETWORK
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
Details
CWE
CWE-798
Status
published
Products (3)
smiths-medical/medfusion_4000_wireless_syringe_infusion_pump
1.1
smiths-medical/medfusion_4000_wireless_syringe_infusion_pump
1.5
smiths-medical/medfusion_4000_wireless_syringe_infusion_pump
1.6