CVE-2018-18563
CRITICALRoche Accu-chek Inform II Firmware - Unrestricted File Upload
Title source: ruleDescription
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.
References (2)
Core 2
Core References
Mitigation, Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01
Third Party Advisory, VDB Entry vdb-entry
x_refsource_bid
http://www.securityfocus.com/bid/105843
Scores
CVSS v3
9.6
EPSS
0.0022
EPSS Percentile
43.9%
Attack Vector
ADJACENT_NETWORK
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Details
CWE
CWE-434
Status
published
Products (5)
roche/accu-chek_inform_ii_firmware
< 03.06.00
roche/coaguchek_pro_ii_firmware
< 04.03.00
roche/coaguchek_xs_plus_firmware
< 03.01.06
roche/coaguchek_xs_pro_firmware
< 03.01.06
roche/cobas_h_232_firmware
< 03.01.03
Published
Nov 20, 2018
Tracked Since