CVE-2018-18565

MEDIUM

Roche Accu-chek Inform II Firmware - Unrestricted File Upload

Title source: rule
STIX 2.1

Description

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.

References (2)

Core 2
Core References
Mitigation, Third Party Advisory, US Government Resource x_refsource_misc
https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01
Third Party Advisory, VDB Entry vdb-entry x_refsource_bid
http://www.securityfocus.com/bid/105843

Scores

CVSS v3 6.8
EPSS 0.0019
EPSS Percentile 40.3%
Attack Vector ADJACENT_NETWORK
CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:C/C:N/I:H/A:N

Details

CWE
CWE-434
Status published
Products (5)
roche/accu-chek_inform_ii_firmware < 03.06.00
roche/coaguchek_pro_ii_firmware < 04.03.00
roche/coaguchek_xs_plus_firmware < 03.01.06
roche/coaguchek_xs_pro_firmware < 03.01.06
roche/cobas_h_232_firmware < 03.01.03
Published Nov 20, 2018
Tracked Since Feb 18, 2026