Description
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Scores
CVSS v3
6.2
EPSS
0.0005
EPSS Percentile
14.7%
Attack Vector
PHYSICAL
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
Details
CWE
CWE-749
Status
published
Products (2)
medtronic/24950_mycarelink_monitor_firmware
medtronic/24952_mycarelink_monitor_firmware
Published
Jul 03, 2018
Tracked Since
Feb 18, 2026