CVE-2019-25719
HIGHDräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Title source: cnaDescription
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
References (3)
Core 3
Core References
Vendor Advisory vendor-advisory
https://static.draeger.com/security
Vendor Advisory vendor-advisory
https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf
Third Party Advisory third-party-advisory
https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling
Scores
CVSS v3
8.6
EPSS
0.0013
EPSS Percentile
3.0%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
yes
Technical Impact
partial
Details
CWE
CWE-924
Status
published
Products (10)
Dräger/Infinity Acute Care System
< VG4.0.3
Dräger/Infinity Acute Care System
< VG4.1.1
Dräger/Infinity Acute Care System
VG4.0.3
Dräger/Infinity Acute Care System
VG4.1.1
Dräger/Infinity Acute Care System
VG4.2
Dräger/Infinity Acute Care System
lower than VG4.0.3
Dräger/Standalone Infinity M540 patient monitor
< VG4.1.1
Dräger/Standalone Infinity M540 patient monitor
VG4.1.1
Dräger/Standalone Infinity M540 patient monitor
VG4.2
Dräger/Standalone Infinity M540 patient monitor
lower than VG4.1.1
Published
Jun 02, 2026
Tracked Since
Jun 02, 2026