CVE-2019-25719

HIGH

Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

Title source: cna
STIX 2.1

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Scores

CVSS v3 8.6
EPSS 0.0013
EPSS Percentile 3.0%
Attack Vector NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

CISA SSVC

Vulnrichment
Exploitation none
Automatable yes
Technical Impact partial

Details

CWE
CWE-924
Status published
Products (10)
Dräger/Infinity Acute Care System < VG4.0.3
Dräger/Infinity Acute Care System < VG4.1.1
Dräger/Infinity Acute Care System VG4.0.3
Dräger/Infinity Acute Care System VG4.1.1
Dräger/Infinity Acute Care System VG4.2
Dräger/Infinity Acute Care System lower than VG4.0.3
Dräger/Standalone Infinity M540 patient monitor < VG4.1.1
Dräger/Standalone Infinity M540 patient monitor VG4.1.1
Dräger/Standalone Infinity M540 patient monitor VG4.2
Dräger/Standalone Infinity M540 patient monitor lower than VG4.1.1
Published Jun 02, 2026
Tracked Since Jun 02, 2026