CVE-2019-5024
HIGHCapsule Technologies SmartLinx Neuron 2 <9.0.3 - Privilege Escalation
Title source: llmDescription
A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.
References (1)
Core 1
Core References
Third Party Advisory x_refsource_confirm
https://talosintelligence.com/vulnerability_reports/TALOS-2019-0785
Scores
CVSS v3
7.6
EPSS
0.0047
EPSS Percentile
37.0%
Attack Vector
PHYSICAL
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Details
CWE
CWE-693
Status
published
Products (1)
capsuletech/smartlinx_neuron_2_firmware
< 9.0.3
Published
Apr 11, 2019
Tracked Since
Feb 18, 2026