CVE-2020-10627

HIGH

Insulet Omnipod Insulin Management System Firmware - Improper Access Control via Wireless RF Communication

Title source: llm
STIX 2.1

Description

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

References (2)

Core 2
Core References
Third Party Advisory, US Government Resource x_refsource_misc
https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01
Vendor Advisory x_refsource_misc
https://www.myomnipod.com/security-bulletins

Scores

CVSS v3 7.3
EPSS 0.0048
EPSS Percentile 37.6%
Attack Vector LOCAL
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L

Details

CWE
CWE-284
Status published
Products (1)
insulet/omnipod_insulin_management_system_firmware
Published Dec 01, 2021
Tracked Since Feb 18, 2026