CVE-2020-12036
HIGHBaxter PrismaFlex and PrisMax < 3.0 - Cleartext Transmission of Sensitive Data to PDMS/EMR
Title source: llmDescription
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
References (1)
Core 1
Core References
Not Applicable x_refsource_misc
https://www.us-cert.gov/ics/advisories/icsma-20-170-01
Scores
CVSS v3
7.5
EPSS
0.0050
EPSS Percentile
38.5%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Details
CWE
CWE-319
Status
published
Products (2)
baxter/prismaflex_firmware
baxter/prismax_firmware
< 3.0
Published
Jun 29, 2020
Tracked Since
Feb 18, 2026