CVE-2020-6962
CRITICALGE Healthcare ApexPro Telemetry Server < 4.2 - Remote Code Execution via Web Configuration Utility
Title source: llmDescription
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X CARESCAPE Central Station (CSCS) Versions 2.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, an input validation vulnerability exists in the web-based system configuration utility that could allow an attacker to obtain arbitrary remote code execution.
References (2)
Core 2
Core References
Third Party Advisory, US Government Resource x_refsource_misc
https://www.us-cert.gov/ics/advisories/icsma-20-023-01
Scores
CVSS v3
10.0
EPSS
0.0493
EPSS Percentile
91.1%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Details
CWE
CWE-20
Status
published
Products (17)
gehealthcare/apexpro_telemetry_server_firmware
4.3
gehealthcare/apexpro_telemetry_server_firmware
< 4.2
gehealthcare/carescape_b450_monitor_firmware
2.0
gehealthcare/carescape_b650_monitor_firmware
1.0
gehealthcare/carescape_b650_monitor_firmware
2.0
gehealthcare/carescape_b850_monitor_firmware
1.0
gehealthcare/carescape_b850_monitor_firmware
2.0
gehealthcare/carescape_central_station_mai700_firmware
1.0
gehealthcare/carescape_central_station_mai700_firmware
2.0
gehealthcare/carescape_central_station_mas700_firmware
1.0
... and 7 more
Published
Jan 24, 2020
Tracked Since