CVE-2021-33885
CRITICALB. Braun SpaceCom2 < 012U000062 - Unauthenticated Insufficient Verification of Data Authenticity
Title source: llmDescription
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
References (2)
Core 2
Core References
Broken Link x_refsource_misc
https://www.bbraunusa.com/en.htm
Exploit, Third Party Advisory x_refsource_misc
https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
Scores
CVSS v3
10.0
EPSS
0.0540
EPSS Percentile
91.6%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N
Details
CWE
CWE-347
Status
published
Products (1)
bbraun/spacecom2
< 012u000062
Published
Aug 25, 2021
Tracked Since
Feb 18, 2026