CVE-2021-37162
CRITICALSwisslog-healthcare Hmi-3 Control Panel Firmware - Buffer Overflow
Title source: ruleDescription
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.
References (4)
Core 4
Core References
Product x_refsource_misc
https://www.swisslog-healthcare.com
Third Party Advisory x_refsource_misc
https://www.armis.com/PwnedPiper
Scores
CVSS v3
9.8
EPSS
0.0756
EPSS Percentile
91.9%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Details
CWE
CWE-120
Status
published
Products (1)
swisslog-healthcare/hmi-3_control_panel_firmware
< 7.2.5.7
Published
Aug 02, 2021
Tracked Since
Feb 18, 2026