CVE-2022-36385
MEDIUMContecHealth CMS8000 Firmware - Unauthenticated Firmware Modification via USB Drive
Title source: llmDescription
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
References (1)
Core 1
Core References
Mitigation, Third Party Advisory, US Government Resource x_refsource_misc
https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01
Scores
CVSS v3
6.8
EPSS
0.0036
EPSS Percentile
27.8%
Attack Vector
PHYSICAL
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
no
Technical Impact
total
Details
CWE
CWE-284
Status
published
Products (1)
contechealth/cms8000_firmware
Published
Sep 13, 2022
Tracked Since