CVE-2022-4992
HIGHDräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering
Title source: cnaDescription
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
References (3)
Core 3
Core References
Vendor Advisory vendor-advisory
https://static.draeger.com/security
Vendor Advisory vendor-advisory
https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf
Third Party Advisory third-party-advisory
https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering
Scores
CVSS v3
8.6
EPSS
0.0016
EPSS Percentile
5.6%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
yes
Technical Impact
partial
Details
CWE
CWE-345
Status
published
Products (5)
Dräger/Infinity Acute Care System
< VG4.0.3
Dräger/Infinity Acute Care System
< VG4.1.1
Dräger/Infinity Acute Care System
< VG4.2
Dräger/Standalone Infinity M540 patient monitor
< VG4.1.1
Dräger/Standalone Infinity M540 patient monitor
< VG4.2