CVE-2023-0888
MEDIUMBbraun Battery-pack SP With Wifi Firmware - Code Injection
Title source: ruleDescription
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Scores
CVSS v3
4.9
EPSS
0.0062
EPSS Percentile
70.1%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
no
Technical Impact
partial
Details
CWE
CWE-95
CWE-94
Status
published
Products (2)
bbraun/battery-pack_sp_with_wifi_firmware
< 053l000092
bbraun/battery-pack_sp_with_wifi_firmware
< 054u000092
Published
Mar 13, 2023
Tracked Since
Feb 18, 2026