B.Braun Battery Pack SP with WiFi Firmware L90/U70 and L92/U92 - Authenticated Eval Injection in Embedded Web Server

Title source: llm
STIX 2.1

Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

References (2)

Core 2

Scores

CVSS v3 4.9
EPSS 0.0060
EPSS Percentile 44.0%
Attack Vector NETWORK
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N

CISA SSVC

Vulnrichment
Exploitation none
Automatable no
Technical Impact partial

Details

CWE
CWE-95 CWE-94
Status published
Products (2)
bbraun/battery-pack_sp_with_wifi_firmware < 053l000092
bbraun/battery-pack_sp_with_wifi_firmware < 054u000092
Published Mar 13, 2023