Description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Scores
CVSS v3
9.3
EPSS
0.0008
EPSS Percentile
22.7%
Attack Vector
LOCAL
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
no
Technical Impact
total
Details
CWE
CWE-494
Status
published
Products (1)
Baxter/Life2000 Ventilation System
06.08.00.00 and prior
Published
Nov 14, 2024
Tracked Since