Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Scores
CVSS v3
9.3
EPSS
0.0009
EPSS Percentile
25.1%
Attack Vector
LOCAL
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
no
Technical Impact
total
Details
CWE
CWE-307
Status
published
Products (1)
Baxter/Life2000 Ventilation System
06.08.00.00 and prior
Published
Nov 14, 2024
Tracked Since
Feb 18, 2026