CVE-2025-0683
MEDIUM EXPLOITEDContec Health CMS8000 Patient Monitor - Info Disclosure
Title source: llmExploitation Summary
CVE-2025-0683 has been observed exploited in the wild (reported by VulnCheck KEV).
Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
References (4)
Core 4
Core References
Various Sources
https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
Third Party Advisory, US Government Resource
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
Third Party Advisory, US Government Resource
https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
Scores
CVSS v3
5.9
EPSS
0.0077
EPSS Percentile
50.5%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
CISA SSVC
Vulnrichment
Exploitation
none
Automatable
no
Technical Impact
partial
Details
VulnCheck KEV
2026-01-20
CWE
CWE-359
Status
published
Products (1)
Contec Health/CMS8000 Patient Monitor
All versions
Published
Jan 30, 2025
Tracked Since
Feb 18, 2026