CVE-2025-14097

HIGH

Multiple Radiometer Products - RCE

Title source: llm
STIX 2.1

Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Scores

CVSS v3 7.2
EPSS 0.0037
EPSS Percentile 58.8%
Attack Vector NETWORK
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

CISA SSVC

Vulnrichment
Exploitation none
Automatable no
Technical Impact total

Details

CWE
CWE-287
Status published
Products (6)
Radiometer Medical Aps/ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems
Radiometer Medical Aps/ABL800 BASIC and ABL800 FLEX Analyzers Application software versions >= 6.20MR2 with Windows 10 operating system
Radiometer Medical Aps/ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems
Radiometer Medical Aps/ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions >= 3.5MR11 with Windows 10 operating system
Radiometer Medical Aps/AQT90 FLEX Analyzers Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems
Radiometer Medical Aps/AQT90 FLEX Analyzers Application software versions >= 8.13 MR2 with Windows 10 operating system
Published Dec 17, 2025
Tracked Since Feb 18, 2026