CVE-2026-0611
CRITICALSpacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Title source: cnaDescription
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
References (3)
Core 3
Core References
Vendor Advisory vendor-advisory
patch
https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf
Product product
https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel
Third Party Advisory third-party-advisory
https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting
Scores
CVSS v3
9.8
EPSS
0.0066
EPSS Percentile
46.7%
Attack Vector
NETWORK
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CISA SSVC
Vulnrichment
Exploitation
poc
Automatable
yes
Technical Impact
total
Details
CWE
CWE-306
Status
published
Products (1)
Spacelabs Healthcare/Sentinel
10.5.0 - 11.6.0
Published
Jun 02, 2026
Tracked Since
Jun 02, 2026