Showing 7 vulnerabilities on this page for Patient Information Center iX (PICiX)

Signals CISA KEV Ransomware Nuclei
Philips vulnerability results
VulnerabilityTitle and contextCVSSEPSSPoCsSignalsSTIX action

Philips Patient Monitoring Devices Exposure of Resource to Wrong Sphere

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges.

CWE-668Sep 11, 2020
CVSS6.8v3.1EPSS0.371%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX

Philips Patient Monitoring Devices Improper Validation of Syntactic Correctness of Input

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.

CWE-1286Sep 11, 2020
CVSS4.3v3.1EPSS0.28%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX

Philips Patient Monitoring Devices Improper Handling of Length Parameter Inconsistency

In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.

CWE-130Sep 11, 2020
CVSS6.5v3.1EPSS0.594%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX

Philips Patient Monitoring Devices Improper Check for Certificate Revocation

In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.

CWE-299Sep 11, 2020
CVSS6.4v3.1EPSS0.368%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX

Philips Patient Monitoring Devices Improper Authentication

In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01, when an actor claims to have a given identity, the software does not prove or insufficiently proves the claim is correct.

CWE-287Sep 11, 2020
CVSS8.8v3.1EPSS0.549%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX

Philips Patient Monitoring Devices Improper Neutralization of Formula Elements in a CSV File

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the software saves user-provided information into a comma-separated value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by spreadsheet software.

CWE-1236Sep 11, 2020
CVSS5.0v3.1EPSS0.586%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX

Philips Patient Monitoring Devices Cross-site Scripting

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output that is then used as a webpage and served to other users. Successful exploitation could lead to unauthorized access to patient data via a read-only web application.

CWE-79Sep 11, 2020
CVSS3.5v3.1EPSS0.653%PoCs0SignalsNot listed in CISA KEVNo known ransomware useNo Nuclei templatesSTIX